ESD Simulator Main Unit Model : ESS-B3011/L1611
Items | ESS-B3011 specifications | ESS-L1611 specifications | |
Polarity | Positive or Negative | ||
Output voltage |
0.2 to 30.0kV +/-5% (30.5kV maximum) 0.2 to 10kV: 0.05kV step 10.0 to 30.5kV: 0.10kV step |
0.2 to 16.0kV +/-5% 0.2 to 10kV: 0.05kV step 10.0 to 16.0kV: 0.10kV step |
|
Repetition cycle |
0.05 to 99.9s (+/- 10%) /Manual 0.05 to 10.0s: 0.05s step 10.0 to 99.9s: 0.10s step |
||
No. of time of discharge | 1 to 999 times /continuous | ||
Discharge mode |
Contact discharge/ Air discharge |
||
Trigger mode | Gun trigger / Main unit trigger | ||
Operation panel |
Indicator: 5x7 dot matrix LED Operation: Push-buttons (Partially lighting) |
||
Radiation mode select switch | Extra / Normal switching function built-in | ||
Discharge detection | Discharge detection function in air-discharge mode | ||
Pre-checking function |
High voltage output checking (by user operation. Not the calibration but just checking) |
||
CR & Gun head checking | CR constant and gun head recognizable (with an indicator to prevent the wrong combination) | ||
IEC level switching function |
Contact discharge mode: 2.0, 4.0, 6.0, 8.0 kV step Air discharge mode: 2.0, 4.0, 8.0, 15.0 kV step |
||
Warning lamp |
Lights up while the HV circuitry is on. Blinks while discharges are in progress. |
||
Charging capacitor and resistance | 150pF+/-10%, 330 ohm +/-10% (built into Discharge Gun GT-30R) | ||
Main unit charging resistance |
10M ohm (53Mohm for combination with Discharge Gun GT-30R) |
||
Power supply/ Power consumption | 100 to 240V AC 50/60 Hz +/-10% 62VA | ||
Included accessories |
Instruction manual |
||
Dimensions and weight | (W)270 X (H)263 X (D)200 mm (projection excluded) Approx. 4.6 kg |
Discharge Gun Model : GT-30R
Items | GT-30R specifications | |
Output voltage | 30.5kV Max | |
Polarity | Positive or Negative | |
Repetition cycle | 0.05s and over Note) Guaranteed value is 0.1s and over (330pF and under) | |
Discharge mode | Contact or Air | |
Charging resistor | Built-in CR unit | |
Charge/Discharge capacitor & Discharge resistor | Built-in CR unit | |
High voltage connector | Composite connector of high voltage and signal line (male) (NoiseKen custom) | |
Operation switch | Discharge trigger | |
LED irradiator | White LED light built-in for irradiating EUT (Focusing point is the tip) | |
Included accessories |
CR unit (150pf-330 ohm) 06-00073A Conical discharge tip 12-00007A Round discharge tip 12-00008A Instruction manual |
|
Dimensions and weight |
(W)83 X (H)217 X (D)229 mm (projection excluded) Approx. 10.8kg (Cable included, connector excluded) |
* Designs and specifications are subject to change without notice.